|
CHEMISTRY 425 Instrumental
methods of analysis
ATOMIC ABSORPTION FLAME PHOTOMETRY OF CADMIUM Start
the cadmium hollow cathode lamp as per instructions and allow about 30 minutes for
warm-up. Adjust the lamp current setting to 5 mA. Use
the stock Cd2+ solution provided to prepare - via dilution in 100 ml volumetric
flasks - standard solutions of 0, 2, 5, 10, 15, 20 and 25 µg/ml Cd2+. Also
dilute your unknown sample to 100 ml and mix thoroughly. NOTE (This note will not be applicable after the repair is
done) The digital readout on the instrument itself is not presently working. Instead, a
voltmeter is connected to the output of the phototube amplifier. Calibration should be set
such that minimum transmittance gives reading 0 V and maximum transmittance will read
5 V. Clearly, one will have to multiply the read potential by 20 to get %
transmittance. The absorbance values will have to be calculated from transmittance. Cadmium vapor
spectrum Ignite
the acetylene-air burner and adjust the flame as instructed. Set the burner elevator and
the gain control at intermediate position and the slit width at 0.5 nm. Set the wavelength
selector at 228.8 nm and adjust for maximum meter deflection. Finally, adjust the gain
control to give a meter deflection of about full scale. (a) Lamp
emission spectrum: While aspirating distilled water into the flame, scan manually from 600
to 200 nm and note the response. Determine ? for the lamp emission peaks and determine
relative intensities. (b) Flame
absorption spectrum: Repeat part (a) but this time while aspirating your standard 25
µg/ml Cd2+ solution into the flame. Readjust the gain control so that the
strongest emission peak remains on scale and repeat any portions of (a) and (b) where the
response was off scale. (c) Selection
of wavelength. Using the results obtained above, calculate the transmittance and the
absorbance at several selected values of wavelength. Explain our choice of 228.8 nm
as the best wavelength for subsequent use. What other values could be employed? Effect of slit
width With
the slit width at 2.0 nm, set the wavelength selector at 228.8 nm and adjust for maximum
response. Readjust the gain control to give approximately full scale deflection. Sweep
manually at low speed the wavelength through the 228.8 nm peak and note the maximum value
and width of the response. Without changing the gain control, repeat at slit width
settings of 2.0, 1.0, 0.5 and 0.20 nm. Interpret the results. Optimum burner
position Set
the slit width at 0.5 nm, adjust the wavelength selector near 228.8 nm for maximum
response, lower the burner elevator as far as it will go, then adjust the gain control to
give about 90% of full scale deflection. Record the response while aspirating distilled
water, then while aspirating your standard 20 µg/ml Cd2+ solution. Without
changing the gain control, raise the elevator by 0.1 inch and repeat, continue to
increments of 0.1 inch until the elevator has been raised 1.0 inch. Plot,
on one graph, response vs. burner position for the water sample and for the
magnesium-containing sample. For each burner position used, calculate transmittance and
absorbance, plot absorbance vs. burner position and choose the optimum position for
subsequent use. Interpret the results you obtained. Empirical
calibration data Set
the slit width at 0.5 nm and - while aspirating distilled water - adjust the gain control
to give precisely full scale response. Aspirate, sequentially, each of your standard
cadmium solutions and your unknown solution and record the response. Take at least three
readings for your unknown. Repeat using a slit width of 0.20 nm, then 1.00 nm. For each
slit width used, plot absorbance vs. concentration of cadmium in the standard solutions.
Determine the concentration of cadmium in your unknown; report the value in ìg/ml of Cd2+
and calculate standard deviation of your reported values. Mention in the discussion any
variations from the original instructions. As
per instructions, extinguish the flame and turn off the instruments. Clean all glassware
and leave the area in a clean, orderly condition. Materials Spectrophotometer: Model Varian Techtron 1200 equipped with a cadmium hollow
cathode lamp. Operating
parameters: Acetylene
pressure 10 psi (do not use an acetylene cylinder with pressure less than 60 psi Air pressure 60
psi Cd hollow cathode
lamp current 3 mA Volumetric
Flasks: Five 100 ml Pipets: One 5 ml,
one 10 ml Stock Solutions: Cd2+ 100 µg/ml Unknown Sample: CdCl2 in a 100 ml volumetric flask
PVZLG PVZAA |